beautypg.com

Fuses & power settings – Carbolite HTR 11/150 User Manual

Page 14

background image


HTR


14

MF30 – 3.07

8.0 F

USES

&

P

OWER

S

ETTINGS

8.1 Fuses

F1-F3: Refer to the circuit diagrams.

F1

Internal
supply
fuses

Fitted if supply cable fitted.
Fitted on board to some types of EMC filter.

on-board and up to 16 Amps:
32mm x 6mm type F
other: GEC Safeclip

F2

Auxiliary
circuit
fuses

Fitted on board to some types of EMC filter.
May be omitted up to 25Amp/phase supply
rating.

2 Amps glass type F
On board: 20mm x 5mm
Other: 32mm x 6mm

F3

Heat Light
fuses

May be omitted up to 25 Amp/phase supply
rating.

2 Amps glass type F
32mm x 6mm

Customer
fuses

Required if no supply cable fitted.
Recommended if cable fitted.

See rating label for amperage;
see table below for fuse rating.

Access to internal fuses is by removal of the back panel of the furnace base or control box.

Model

phases

Volts

Supply Fuse
Rating

Volts

Supply Fuse
Rating

HTR 11/75

1-phase

200-240

10A

110-120

16A

HTR 11/150

1-phase

220-240

12.5A

110-120

25A

other models or voltages: check the rating label for details of the supply.