beautypg.com

Carbolite HTR 11/150 User Manual

Page 10

background image


HTR


10

MF30 – 3.07

6.0 F

AULT

A

NALYSIS

A. Furnace Does Not Heat Up

1.

The HEAT light
is ON

 The heating

element has failed

 Check also that the SSR is working

correctly

2.

The HEAT light
is OFF

 The controller

shows a very high
temperature
or a
code such as S.br

 The thermocouple has broken or has a

wiring fault

 The controller

shows a low
temperature

 The door switch(es) (if fitted) may be

faulty or need adjustment

 The contactor (if fitted) may be faulty

 The SSR could be failing to switch on due

to internal failure, faulty logic wiring from
the controller, or faulty controller

 There are no lights

glowing on the
controller

 Check the supply fuses and any fuses in the

furnace control compartment

 The controller may be faulty or not

receiving a supply due to a faulty switch or
a wiring fault

B. Furnace Overheats

1.

The HEAT light
goes OFF with
the instrument
switch

 The controller

shows a very high
temperature

 The controller is faulty

 The controller

shows a low
temperature

 The thermocouple may have been shorted

out or may have been moved out of the
heating chamber

 The thermocouple may be mounted the

wrong way round

 The controller may be faulty

2.

The HEAT light
does not go off
with the
instrument
switch

 The SSR has failed

“ON”

 Check for an accidental wiring fault which

could have overloaded the SSR

This manual is related to the following products: