Eneral, Eter, Pecifications – Red Lion PAXLSG User Manual
Page 3: Ccessories

1. DISPLAY: 3 1/2-digit, 0.56" (14.2 mm) high, 7-segment red LED, (-) minus
sign displayed when voltage is negative. Decimal points inserted before 1st,
2nd, or 3rd least significant digits by DIP switch selection.
2. OVER-RANGE INDICATION: Indicated by blanking 3 least significant
digits.
3. POWER:
AC Power: 85 to 250 VAC, 50/60 HZ, 6 VA
Isolation: 2300 Vrms for 1 min. to all inputs.
4. INPUT SIGNAL: Single-ended or differential input, ±2.0 V max. Gain
(Sensitivity) is adjustable from 200 Units of Numerical Readout/millivolt input
(gives full scale readout of 1999 at 10 mV input), to less than 1 Unit of
Numerical Readout/mV (gives full scale readout of 1999 at 2.0 V input).
Maximum common mode voltage swing with respect to signal ground, 0 to 7 V.
Note: Absolute maximum voltage that can be applied between the two input
terminals or between input and signal common is 50 VDC.
5. INPUT IMPEDANCE: 100 M
6. LINEARITY: ±(0.05% ±1 digit)
7. LOW FREQUENCY NOISE REJECTION:
Normal Mode Rejection: 84 dB @ 50/60 Hz
Common Mode Rejection: 50 dB with respect to excitation common;
110 dB with respect to earth ground.
8. RESPONSE TIME: 2.0 seconds to settle from step input.
9. READING RATE: 2.5 updated readings/second, nominal.
10. EXCITATION SUPPLY:
Jumper Selectable: 5 VDC @ 60 mA max., ±2%
10 VDC @ 120 mA max., ±2%
Temperature coefficient (ratio metric): 20 ppm/°C max.
11. ENVIRONMENTAL CONDITIONS:
Operating Temperature: 0° to 60°C
Storage Temperature: -40° to 80°C
Operating and Storage Humidity: 85% max. relative humidity (non-
condensing)
Span Temperature Coeff.: 100 PPM/°C
Offset Temperature Coeff.: 100 PPM/°C
Vibration According to IEC 68-2-6: Operational 5 to 150 Hz, in X, Y, Z
direction for 1.5 hours, 2g.
Shock According to IEC 68-2-27: Operational 30 g, 11 msec in 3 directions.
Altitude: Up to 2000 meters
12. CERTIFICATIONS AND COMPLIANCES:
SAFETY
UL Recognized Component, File # E179259, UL61010A-1, CSA C22.2 No. 61010-1
Recognized to U.S. and Canadian requirements under the Component
Recognition Program of Underwriters Laboratories, Inc.
UL Listed, File # E137808, UL508, CSA C22.2 No. 14-M95
LISTED by Und. Lab. Inc. to U.S. and Canadian safety standards
Type 4X Enclosure rating (Face only), UL50
IECEE CB Scheme Test Certificate # UL/8843A/UL
CB Scheme Test Report # 04ME11209-20041018
Issued by Underwriters Laboratories, Inc.
IEC 61010-1, EN 61010-1: Safety requirements for electrical equipment
for measurement, control, and laboratory use, Part 1.
IP65 Enclosure rating (Face only), IEC 529
IP20 Enclosure rating (Rear of unit), IEC 529
ELECTROMAGNETIC COMPATIBILITY
Emissions and Immunity to EN
61326: Electrical
Equipment for
Measurement, Control and Laboratory use.
Notes:
1. Criterion A: Normal operation within specified limits.
2. Criterion B: Temporary loss of performance from which the unit self-
recovers.
13. CONNECTIONS: High compression cage-clamp terminal block
Wire Strip Length: 0.3" (7.5 mm)
Wire Gage: 30-14 AWG copper wire
Torque: 4.5 inch-lbs (0.51 N-m) max.
14. CONSTRUCTION: This unit is rated for NEMA 4X/IP65 outdoor use.
IP20 Touch safe. Installation Category II, Pollution Degree 2. One piece
bezel/case. Flame resistant. Panel gasket and mounting clip included.
15. WEIGHT: 0.65 lbs (0.24 kg)
3
G
ENERAL
M
ETER
S
PECIFICATIONS
A
CCESSORIES
UNITS LABEL KIT (PAXLBK)
Each meter has a units indicator with backlighting that can be customized
using the Units Label Kit. The backlight is controlled by a DIP switch.
Immunity to Industrial Locations:
Electrostatic discharge
EN 61000-4-2
Criterion A
4 kV contact discharge
8 kV air discharge
Electromagnetic RF fields
EN 61000-4-3
Criterion B
10 V/m
2 kV L&N-E power
Fast transients (burst)
EN 61000-4-4
Criterion B
2 kV power
2 kV signal
Surge
EN 61000-4-5
Criterion A
1 kV L-L,
1 kV signal
RF conducted interference
EN 61000-4-6
Criterion A
3 V/rms
Power frequency magnetic fields EN 61000-4-8
Criterion A
30 A/m
Voltage dip/interruptions
EN 61000-4-11
Criterion A
0.5 cycle
Emissions:
Emissions
EN 55011
Class B