ApplicAtion

instructions

ADDITIONAL OR DIFFERENT BODY OR LIMB RESTRAINTS MAY

BE NEEDED (See Posey Catalog):

• If the patient pulls violently against the bed straps;

• To reduce the risk of the patient getting access to the line/

wound/tube site;

• To prevent the patient from flailing or bucking up and down and

causing self-injury.

• Check the patient regularly to ensure that:

Circulation is not impaired. Serious injury may occur if the cuffs

restrict circulation.

Cuffs are secure. Death or serious injury to the patient or

others may occur if the patient can remove the cuffs.

MONITORING

• Posey recommends constant direct supervision for patients

deemed to be at risk of injury to themselves or others. For times

when direct supervision is not possible, monitor by line of sight or

by a video/audio device.

• NEVER allow a patient to have access to any tool, utensil, or

object that might be used to unlock or damage cuffs.

• Be aware that a sudden mood swing may cause agitated or

aggressive behavior. Contact the medical team AT ONCE if this

occurs. No level of monitoring may avoid the risk of serious injury

to highly agitated or aggressive patients.

BED SAFETY

• ALWAYS use Hospital Bed Safety

Workgroup (HBSW) (http://www.fda.gov/

cdrh/beds/modguide.html) compliant side

rails in the UP position and fill ALL gaps to

reduce the risk of entrapment.

• Use side rail covers and gap protectors to help prevent the

patient’s body from going under, around, through or between the

side rails. A failure to do so may result in serious injury or death

if a patient becomes suspended or entrapped. Posey offers a full

range of side rail pads and/or gap protectors to cover gaps.

ApplicAtion instructions:

• Before each use, check cuffs and straps for cracks, tears, and/or

excessive wear or stretch; cracked or broken buckles or locks;

and/or that hook and loop adheres securely, as these may allow

patient to remove cuff. Discard if device is damaged.

• If you have any questions about patient safety, ask the

appropriate medical authority for alternatives.

Following these steps to apply device (repeat steps 1-4 on the other limb):

1. Loop the connecting strap through the U-bar on the cuff.

2. Wrap the strap around a movable part of

the bed frame, out of the patient’s reach.

Ensure that the strap is wrapped around

the frame at least once before passing

the end of the strap through the buckle

(fig. 1).

3. Pull the strap through the roller buckle.

4. Thread excess strap through the bar on the strap.

MONITOR PER FACILITY POLICY. Check to ensure that:

• Straps cannot slide in any direction, tighten or loosen if the

patient pulls on them, or if the bed is adjusted;

• Cuffs are attached in a way that the patient is not able to use his

or her teeth or otherwise remove the device;

• Cuffs are intact, and not torn or damaged. DO NOT allow

patients to ingest product material.

Applicable Products:

REF

2330, 2346, 2355

DEscription oF proDuct: Connecting Straps for use only in conjunction with Posey Leather, Synthetic Leather or E.D. Security Cuffs.

rx onlY

non-locKinG connEctinG strAps

© 2009 Posey Company. All rights reserved.

I9254 062109

2346

2355

2330

posEY non-locKinG connEctinG strAps

REF

2330 Leather strap w/roller buckle 18” (46 cm), 24” (61 cm), 36” (91 cm), 48” (122 cm)

REF

2346 Biothane strap w/roller buckle 18” (46 cm), 24” (61 cm), 36” (91 cm), 48” (122 cm), 60” (152 cm), 72” (183 cm)

REF

2355 Synthetic Leather strap w/roller buckle 18” (46 cm), 24” (61 cm), 36” (91 cm), 48” (122 cm), 60” (152 cm), 72” (183 cm)

FiG. 1

ADVErsE rEActions

• Severe emotional, psychological, or physical problems may occur: if the applied device is uncomfortable;

or if it severely limits movement. If the patient is likely to cause injury to him/herself or others, get help

from a qualified medical authority to find an alternate intervention or method of restraint.

inDicAtions For usE:

• Patients assessed as being in extreme danger of injury to themselves or to others.

• Patients assessed to be at risk of disrupting life-saving treatments (e.g., chronic tube pulling).

• Patients assessed to be at risk of line pulling, which may prevent monitoring of vital signs.

• Patients whose picking, pulling, scratching, or peeling exacerbates a skin condition, causes self-injury, or

compromises wound site integrity.

contrAinDicAtions:

NEVER use on a patient:

With a dislocation or fracture on the restrained limb; or

If an IV or wound site could be compromised by the device.

See the Posey Catalog for other options for such a patient.

DO NOT use this device on a patient who is or becomes: suicidal; highly aggressive or combative;

self-destructive; or deemed to be an immediate risk to others, UNLESS the patient is under constant

supervision.

• NEVER use the connecting strap directly on a patient.

ADDitionAl sAFEtY AnD lAunDErinG