Limits of
temperature at flange:
Version
Minimum flange temperature
Maximum flange temperature
(1)+(2) of the
type code
Standard
version
Special version
with marking
„2.4610“ at the
Metaglas ring
Without high
temperature
distance piece
With high
temperature
distance piece
V96 ... FFKM
- 30°C (- 22°F)
- 60°C (-76°F) +130°C (+266°F) +250°C (+482°F)
V96 ... K2035
- 30°C (- 22°F)
- 60°C (-76°F) +130°C (+266°F) +210°C (+410°F)
V96 ... FPM
- 30°C (- 22°F)
- 60°C (-76°F) +130°C (+266°F) +200°C (+392°F)
V96 ... FEP
- 30°C (- 22°F)
- 60°C (-76°F) +130°C (+266°F) +200°C (+392°F)
WS LPTFE
- 40°C (- 40°F)
–––
+130°C (+266°F) +150°C (+302°F)
WS PTFE
- 40°C (- 40°F)
–––
+130°C (+266°F) +150°C (+302°F)
WS SS PTFE
- 20°C (- 4°F)
–––
+130°C (+266°F) +150°C (+302°F)
WS SS PP
- 20°C (- 4°F)
–––
+100°C (+212°F) +100°C (+212°F)
9 Parameter check list BM 700
Menu item changed on Version:
Fct. Configuration parameters (extract)
3.1.1 Tank height
3.1.2 Block distance
3.1.3 Antenna
3.1.4 Antenna extension
3.1.5 Distance piece
3.1.6 Stillwell / diameter
3.1.7 Reference offset
3.1.8 Tank bottom
3.3.1 Current output, function offset
3.3.2 Current output range/error
3.3.3 Min. current scale
3.3.4 Max. current scale
3.5.2 Empty spectrum
3.5.3 Time constant
3.5.4 Tracking speed
3.5.5 Multiple reflections (yes/no)
3.5.6 Block distance ident (yes/no)
3.5.7 Function FTB
3.5.8 Epsilon R
3.5.9 Tank type