Test and troubleshooting, Calibration, 1pa100 – Pacific Laser Equipment 1PA-100 User Manual
Page 7
Amplifier 1ch100
User manual
1PA100
October 10, 2004
Release: 1.30
2002 by
Pacific Laser Equipment
Test and troubleshooting
High voltage output
Attention:
The HV output can reach +150V. This is a dangerous voltage.
Test in RC mode. At 5V analog input the indicator should read MAX.
Fuse
Check the main fuse inside the unit. When replacing it, use slow-blow 100mA.
The fuse is located inside the box and access to it is fairly difficult.
AC Power
Make sure that the AC power used is always 120V.
Calibration
Adjusted during manufacturing RC mode with the front panel potentiometer fully turned
counterclockwise and 0V analog input, the HV output is 0V.
At 5V analog input the HV output is 150V DC.
4.
Calibration
PLEquipment, 3941 S. Bristol St. Unit E, # 122, Santa Ana, CA. 92704,
800 984-3482, Fax: 509-355-5155, [email protected]