Section 8 troubleshooting of the model iaq10 – TSI IAQ10 User Manual
Page 15

13
SECTION 8
Troubleshooting of the Model IAQ10
This table lists the symptoms, possible
causes, and recommended solutions for the
Model IAQ10.
Symptom
Possible
Causes
Corrective
Action(s)
No display Unit not
switched on
Switch on the
unit
Low or dead
batteries
Replace the
batteries
Dirty
battery
contacts
Clean the
battery
contacts
Batteries
installed
incorrectly
Refer to
battery
illustration
inside battery
cover
BAT is
blinking
Dirty battery
contacts
Batteries are
low
Clean the
battery
contacts
Replace the
batteries
Cal err
Error in field
calibration
Perform
calibration
again