Input, Measurement function – Atec Yokogawa-cw240 User Manual

Page 2

background image

2

All Rights Reserved. Copyright © 2004, Yokogawa M&C Corporation

Input

Item

Voltage (V)

Input type

Resistive potential division

Rated Value (Range)

150.0/300.0/600.0/1000V

96036 (2A)

200.0/500.0mA/1.000/2.000A

96033 (50A)

5.000/10.00/20.00/50.00 A

96030 (200A)

20.00/50.00/100.0/200.0 A

96031 (500A)

50.00/100.0/200.0/500.0 A

96032 (1000A)

200.0/500.0 A/1.000 kA

96034 3000A range

300.0/750.0 A/1.500/3.000 kA

2000A range

200.0/500.0 A/1.000/2.000 kA

   

1000A range

100.0/200.0/500.0 A/1.000 kA

96035 3000A range

300.0/750.0 A/1.500/3.000 kA

    

300A range

30.00/75.00/150.0/300.0 A

Phase/Wiring Measurement

Number on measurement system

Input resistance

Approximately 1.3MΩ

Maximum allowed

1000Vrms

96036 (2A)

20Arms

continuous input

96033 (50A)

130Arms

96030 (200A)

250Arms

96031 (500A)

625Arms

96032 (1000A)

700Arms

96034 3000A range

2400Arms

2000A range

2400Arms

   

1000A range

1200Arms

96035 3000A range

3600Arms

    

300A range

360Arms

AD conversion

Common Voltage
1P2W: 4 system load/ 1P3W: 2 system load/ 3P3W2i: 2 system load

Approximately 100kΩ(CW240)

Simultaneous inversion of Voltage/Current input, PPL synchronism

Current (A)

Clamp detection

1P2W/1P3W/1P3W3i(Nutral line current)/3P3W2i(2-wattmeter method)/3P3W3i(3-
wattmeter method),3P4W/3P4W4i(Nutral line current), SCOTT wiring (1P3W + 3P3W)

Measurement function

Parameter

Voltage

Method
Frequency range
Crest factor
Accuracy

±0.2%rdg.±0.1%rng. 96030, 96031, 96033, 96036 ±0.6%rdg.±0.4%rng.

96032, 96034, 96035

±1.0%rdg.±0.8%rng.

Power factor effects

96030 ±1.0%rng.(45 to 65Hz,Power factor = ±0.5)

Except 96030

±2.0%rng.(45 to 65Hz,Power factor = ±0.5)

Effect of reactive factor

96030 ±1.0%rng.(45 to 65Hz,Power factor = ±0.5)

Except 96030

±2.0%rng.(45 to 65Hz,Reactive factor= ±0.5)

Valid input range

Display Range

Temperature coefficient

±0.03%rng/℃

Display update period

rdg: reading   rng: range

Current/Active Power/Reactive Power1(Use reactive-power meter method)

Voltage/Current: 0.4 to 130% for each range (Zero suppress less than 0.4%)
Electric Power(Active, Reactive, Apparent): 0 to 130% for each range (Zero suppress if less
than 0.17% of range rating)
Harmonics Level: 0 to 130% for each range

±0.05%rng/℃

Approximately 0.5 second

Digital sampling
45Hz to 65Hz (Select measurement element from U1, U2 or U3)
3 for rated input (1.8 for 1000 range)

5 to 110% for each range (Max 100% for 1000 range)