Alarm information
E8.1
Fault diagnosis related to E 8.1
The empty adjustment data has not been transferred
Repeat empty adjustment,
see page 11
Alarm E 8.1 extinguishes
O.K.
?
Auswertgerät defekt
E8.2
The full adjustment data has not been transferred
Repeat full adjustment,
see page 11
Alarm E 8.2 extinguishes
O.K.
?
Auswertgerät defekt
E8.3
The integration time data has not been transferred
Repeat programming of integration
time, see page 12
Alarm E 8.3 extinguishes
O.K.
?
Auswertgerät defekt
E8.4
The password data has not been transferred
Repeat activation or adjustment of
password, see page 13
Alarm E 8.4 extinguishes
O.K.
?
Auswertgerät defekt
E9
Try a new set-up, see page 10
Alarm E 9 extinguishes
O.K.
?
Auswertgerät defekt
Fault (data loss) in the EEPROM-memory
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Signal conditioning
instrument defect
Signal conditioning
instrument defect
Signal conditioning
instrument defect
Signal conditioning
instrument defect
Signal conditioning
instrument defect