Metrologic Instruments Genesis MS7580 User Manual
Page 31

27
T
ROUBLESHOOTING
G
UIDE
Symptoms
Possible Causes
Solution
The unit scans
but the data is
not correct.
The scanner and host
may not be configured
for the same interface
parameters.
Check that the scanner and the
host are configured for the same
interface parameters.
The following item is only relevant for an RS232 interface.
The unit powers
up OK and scans
OK but does not
communicate
properly with the
host.
The com port at the
host is not working or
not configured
properly.
Check to make sure that the baud
rate and parity of the scanner and
the communication port match
and the program is looking for
“RS232” data.
The unit powers
up OK and scans
OK but does not
communicate
properly with the
host.
The cable is not
connected to the
correct com port.
Check to make sure that the cable
is connected to the correct com
port.
Characters are
being dropped.
Inter-character delay
needs to be added to
the transmitted output.
Add some inter-character delay to
the transmitted output by using
the Configuration Guides
(
MLPN
00-02544 and 00-05252).
The following item is only relevant for a USB interface.
The USB host may not
be active.
The MS7580 will not
operate from hub/host
power without host
communication.
Turn on the host device.
No LEDs, beep
or Illumination
The unit is not
receiving sufficient
power to support
operation.
Refer to the electrical
specifications on page 29.
Verify adequate power is being
supplied to the scanner.