Compact™ IO Isolated Thermocouple Module

User’s Manual Pub. 0300244-01 Rev. A

Thermocouple Type
N (-200

°C to 1300°C)

+/- 1 degrees C maximum @ 25 degrees C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 1.5 degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

Thermocouple Type
N (-210

°C to -200°C)

+/- 1.2 degrees C maximum @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 1.8 degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

Thermocouple Type
T (-230

°C to 400°C)

+/- 1 degrees C maximum @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 1.5 degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

Thermocouple Type
T (-270

°C to -230°C)

+/- 5.4 degrees C maximum @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 7.0 degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

Thermocouple Type
K (-225

°C to 1370°C)

+/- 1 degrees C maximum @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 1.5 degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

Thermocouple Type
K (-270

°C to -225°C)

+/- 7.5 degrees C maximum @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 10 degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

Thermocouple Type
E (-210

°C to 1000°C)

+/- 0.5 degrees C maximum @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 0.8 degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

Thermocouple Type
E (-270

°C to -210°C)

+/- 4.2 degrees C maximum @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 6.3 degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

Thermocouple Types
S and R

+/- 1.7 degrees C maximum @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 2.6 degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

Thermocouple Type C

+/- 1.8 degrees C maximum @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 3.5 degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

Thermocouple Type B

+/- 3.0 degrees C maximum @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 4.5 degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

Thermocouple Type L

+/- 0.6 (TBD) degrees C maximum @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 0.9 (TBD) degrees C maximum @ 0 – 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

CJC Sensor range

0 to 85 degrees C

CJC Sensor accuracy

+/- 0.2 degrees C maximum

CJC Measurement accuracy

+/- 0.1 degrees C maximum

CJC Profile accuracy

+/- 0.7 degrees C maximum

CJC Total accuracy
(sum of above)

+/- 1.0 degrees C maximum

Millivolt Inputs

+/- 15 uV maximum for +/- 50 mV inputs @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 25 uV maximum for +/- 50 mV inputs @ 0 - 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 20 uV maximum for +/- 100 mV inputs @ 25°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz filters.

+/- 30 uV maximum for +/- 100 mV inputs @ 0 - 60°C for

4.17 Hz, 10 Hz, 16.7 Hz, and 19.6 Hz

filters.

Note: Accuracy is dependent on the ADC filter selection, data format, and input noise.

Repeatability (at 25

°C)

8

4.17 Hz and 10Hz filter

Thermocouple Types
J, L and N

+/- 0.1°C

Thermocouple Types
N (-110

°C to 1300°C)

+/- 0.1°C

Thermocouple Types
N (-210

°C to -110°C)

+/- 0.25°C

Thermocouple Types
T (-170

°C to 400°C)

+/- 0.1°C

Thermocouple Types
T (-270

°C to -170°C)

+/- 1.5°C

Thermocouple Type
K (-170

°C to 1370°C)

+/- 0.1°C

Thermocouple Type
K (-270

°C to -170°C)

+/- 2.0°C

Thermocouple Type
E (-220

°C to 1000°C)

+/- 0.1°C

Thermocouple Type
E (-270

°C to -220°C)

+/- 1.0°C

Thermocouple Types
S and R

+/- 0.4°C

8

Repeatability for:

16.7 Hz filter is 2X the repeatability of the 4.17Hz filter
19.6 Hz filter is 2X the repeatability of the 4.17Hz filter
62 Hz filter is 4X the repeatability of the 4.17Hz filter
470 Hz filter is 10X the repeatability of the 4.17Hz filter