Chapter 6: Maintaining Your Enclosure
Product Service 1-800-522-7658
15
Fluorescent Light Replacement
1. Disconnect power.
2. Locate the two Phillips machine screws on top of the light and remove.
3. Gently lift up the light assembly and rotate it upside down to rest on top of
the enclosure. Use caution when lifting and removing light assembly to
prevent damage to UV closure panel sensor wire, sensor wire is attached
between light assembly and enclosure.
4. Rotate and remove the old fluorescent lamp.
5. Reinstall the new fluorescent lamp, light and screws in reverse order. Use
caution not to pinch or damage UV closure panel sensor wire.
6. Switch unit on and try the new fluorescent lamp.